Assembly
42 Instructions for use Oxylog 3000 plus SW 1.n
2 Connect the compressed gas hose for O2 (A) to
Oxylog 3000 plus.
3 Connect the O
2 hose to the
pressure reducer (B).
4 Rotate the cylinder valve (C) slowly and open
fully.
O2 supply from a central gas
supply system
1 Connect the compressed gas hose for O2 (A) to
Oxylog 3000 plus.
2 Connect the gas probe (B) to the O
2 terminal
unit until it has properly engaged and the supply
of O
2 is assured.
018
WARNING
Risk of ventilator malfunction
Do not install flow control valves or
flowmeters in the gas supply to
Oxylog 3000 plus.
WARNING
Always check the O2 pressure of cylinder
before use, to prevent insufficient oxygen
supply during use.
O
2
A
B
C
019
A
B