Assembly
36 Instructions for use Oxylog 3000 plus SW 1.n
WARNING
Do not use any damaged parts or
accessories.
Damaged or deformed parts must be
replaced.
WARNING
Electrical connections to equipment, which
are not listed in these instructions for use,
should only be made following consultation
with the respective manufacturers.
Equipment malfunction may result as well as
risk of patient injury.
WARNING
All equipment connected to Oxylog 3000 plus
must comply with IEC 60601-1-2.
WARNING
Risk of CO2 rebreathing
Do not combine parts of different breathing
circuits, especially for pediatric applications.
WARNING
Always use the angled connector of the
breathing circuit.
If the angled connector is not used, the
minute volume may be measured incorrectly.
CAUTION
Do not use electrically conductive hoses.
This may endanger the user and may cause dam-
age to the device during defibrillation.
Risk of electric shock.