Maintenance
Dräger Polytron 8100 31
6.3 Troubleshooting
6.3.1 Fault reference
Fault number Cause Remedy
#001
#003
#004
#005
#011 - #014
#020 - #024
#043
#060
#067
Serious instrument fault,
various causes.
Have the instrument checked by DrägerService
®1
.
#002
#025 - #027
#030 - #034
#041
#042
#044
#050
#052
#081
#085
Serious data error in the instrument,
various causes.
Reset the instrument to factory default settings (see Section 7
on Page 34).
If this fault occurs again:
Have the instrument checked by DrägerService.
#010 4 to 20 mA interface cable not connected. Disconnect the power supply, connect the 4 to 20 mA interface
and reconnect the power supply.
If operated as a stand-alone instrument without a central
controller, wire pin 3 to pin 4 (see Section 3.5 on Page 9).
#045 Instrument not detecting any sensor. Check connections.
If this fault occurs again:
Have the instrument checked by DrägerService.
#051
#054
Faulty zero calibration. Perform zero calibration (see Section 6.2.1 on Page 29).
#055 Faulty span calibration. Perform span calibration (see Section 6.2.2 on Page 29).
#076
#080 - #084
Instrument fault. Check electrical connections.
If this fault occurs again:
Have the instrument checked by DrägerService.
#090 Sensor lock function is activated. A sensor
with a different part number has been
connected.
Use a sensor with the same part number or disable the sensor
lock function (see Section 5.9.3 on Page 25).
#105 Instrument fault. Have the instrument checked by DrägerService.
#137 Instrument fault. Cycle power.
If this fault occurs again:
Have the instrument checked by DrägerService.
1 DrägerService
®
is a registered trademark of Dräger.