32 Dräger Polytron 8100
Maintenance
6.3.2 Warning reference
Warning
number
Cause Remedy
#101 Data-logger in Stack mode is 100 % full and
is not logging any more data.
Download the data, clear the data-logger.
#102 Data-logger in Stack mode is 90 % full. Download the data as soon as possible and clear the data-
logger.
#103
#106
Data error in the instrument. Some dongle
functions like data-logger, sensor test, etc.
may no longer be available.
Reset the instrument to factory default settings (see Section 7
on Page 34).
If this error occurs again:
Have the instrument checked by DrägerService.
#104 Date or time setting not valid. Set date and time (see Section 5.7.3 on Page 20).
#107 Battery for data memory empty. Have the instrument checked by DrägerService.
#110
#111
#112
SW dongle has been removed without being
deactivated.
Deactivate SW dongle (see Section 5.10 on Page 27).
#113 Alarms inhibited. Enable alarms (see Section 5.7.1 on Page 19).
#164
#165
Sensor warm-up phase has not ended.
Increased measurement error must be
expected.
Wait until the sensor has warmed up. Do not calibrate before
sensor is fully warmed-up.
#167
#170
Calibration interval expired. Recalibrate the instrument (see Section 6.2 on Page 28).