Technical data
Dräger Polytron 8100 37
10 Technical data
10.1 Approvals
10.1.1 ATEX, IECEx, UL, CSA
See printout of approval label inside the shipping box of the
instrument.
10.1.2 DIN EN 50104 for DrägerSensor O
2
(6809720)
Sensor Principle
The DrägerSensor O
2
(6809720) is an electrochemical two-
electrode sensor for measuring Oxygen (O
2
) in ambient air.
CAUTION
Specifications and restrictions in the Instructions for
Use and/or data sheets for the sensors used must be
observed. For SIL applications the Dräger Polytron
8X00 Safety Manual must be observed.
The measuring range and performance
characteristics are dependent on the installed sensor
(see the Instructions for Use and/or data sheet for the
installed sensors).
Reaction at electrodes
Measuring electrode
O
2
+ 2 H
2
O + 4 e
=> 4 OH
Counter electrode
2 Pb => 2 Pb
2
+ + 4 e
Environmental parameters
Pressure 20.7 to 38.4 in. Hg (700 to1300 hPa)
Humidity 10 to 95 %RH, non-condensing
Temperature -5 to +40 °C, short term -20 to +55 °C
Storage humidity 30 to 70 %RH, non-condensing
Storage temperature 0 to +40 °C
!
D
Polytron 8000
ETR 0401
Dräger Safety 23560 Lübeck, Germany
II 2G
II 2D
WARNING: Do not open when energized
I
S
F
S
I
E
A
D
L
C
9N54
0158
Class I, Div 1, Groups A,B,C,D | Class II, Div 1, Groups E,F,G
Class I, Zone 1, Group IIC | T-Code T6/T4 | Type 4X
Supply: 10...30 VDC, 0.1...0.35 A | Relay: 5 A, 30 VDC or 230 VAC
CAUTION: Do not open cover. Opening
cover with circuits alive provokes risk of
Ignition of Hazardous Atmospheres.
W
ARNING: To reduce the risk of Ignition
of Hazardous Atmospheres, the conduit
must be sealed within 18’’ of the enclosure.
WARNING: Read Manual before operating.
Gas Detector for Use in Hazardous Locations as to
Fire, Electrical Shock and Explosion Hazards only
Part No: 4544404
Serial No. XXXX-9999
Ex d [ia] IIC T6/T4 Gb
Ex tb IIIC T80/130°C Db
-40°C Ta +40/+80°C
PTB 11 ATEX 1005X
IECEx PTB 11.0005X
IP6x | P 5 W | UL only: -40°C °C Ta +70
Sample
Response time
t
0...20
< 12 seconds
t
0...90
< 30 seconds
NOTICE
At temperatures below -5 °C, the response time may
increase.
Calibration
Flow rate 0.5 L/min
Zero gas Nitrogen (99.9 Vol% N
2
)
Target gas Oxygen / Nitrogen gas mixture
Warm-up time
Operation < 15 minutes
Calibration <
2 hours
i