EFW5L_IFU REV NEW
Guidance and manufacturer’s declarationelectromagnetic emissions
EFW5L is intended for use in the electromagnetic environment specified below. The customer or
the user of EFW5L should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
RF emissions
CISPR 11
Group 1
EFW5L uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class A
Harmonic
emissions IEC
61000-3-2
Complies
Voltage
fluctuations
and flicker
emissions IEC
61000-3-3
Complies
Table 2: Guidance and Manufacturer's Declaration Electromagnetic Emissions
the user of EFW5L should assure that it is used in such an environment.
Immunity
test
IEC 60601 test
level
Compliance
level
Electromagnetic environment guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete, or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast
transient/
burst IEC
61000-4-4
±2 kV for power
supply lines
±1 kV for
input/output lines
±2 kV for power
supply lines
Not applicable
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-
5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be that of a
typical commercial or hospital environment.