EFW5L_IFU REV NEW
Symbol Symbol Title
(Symbol Description)
Standard Title Reference
IC Certification Mark
Certifies that EFW5L has been tested and
found to be in compliance with Industry
Canada (IC) standards.
Information Technology
Equipment
(including
Digital Apparatus)
IC ICES-003
Table 1: Labels and Symbols
1.4 Warnings and Precautions
EFW5L is designed for use by Healthcare professionals.
EFW5L is designed for transport and storage in indoor ambient temperature and
humidity conditions: -18ºC to 60ºC (0ºF to 140ºF), 15% to 90% RH, 70 to 106 kPa
atmospheric pressure.
EFW5L is designed for use in indoor ambient temperature and humidity conditions: 16°C
to 27°C (60°F to 80°F), 45% to 60% RH, 70 to 106 kPa atmospheric pressure.
EFW5L is not designed for use, transport, or storage in an oxygen rich environment.
EFW5L is not designed for use, transport, or storage in the presence of flammable
anesthetics
Position EFW5L to allow for easy removal of power cord from power outlet. Power cord
must be disconnected from outlet to isolate EFW5L from external power supply.
Warning: No direct patient contact! Do not place a patient (e.g. an infant) or any part of a
patient into, on, or against EFW5L or into direct contact with the warming basin. Contact
may cause thermal injury.
Warning: To avoid risk of electric shock, this equipment must only be connected to a
supply main with protective earth.
Warning: Risk of Fire: Replace fuse as marked. Refer to Rating Label and Section 9.3
for Fuse information.
Warning: Do not sit or stand on EFW5L.
Warning: Do not transport at an incline of greater than 5°.
Warning: Do not transport while unit is draped and surgical solution is present.
Warning: No modification of this equipment is allowed.
Warning: Grounding reliability can only be achieved when this equipment is connected to
an equivalent receptacle marked “Hospital Only” or “Hospital Grade.”
Warning: Do not operate without approved Ecolab drape and recommended fluid
volumes.
Warning: Power must be off until minimum fluid level is poured into drape.
Caution: Electric shock hazard, do not remove protective panel. Refer servicing to
qualified service personnel.
Caution: Make sure that the warmer is turned off before unplugging the power cord.
Warning: Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are operating
normally.