• Warning: Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment
and result in improper operation.
• Warning: Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30 cm (12 inches)
to any part of the EFW5L including cables specified by the manufacturer. Otherwise,
degradation of the performance of this equipment could result.
• Note: The EMISSIONS characteristics of this equipment make it suitable for use in
industrial areas and hospitals (CISPR 11 class A). If it is used in a residential
environment (for which CISPR 11 class B is normally required) this equipment might not
offer adequate protection to radio-frequency communication services. The user might
need to take mitigation measures, such as relocating or re-orienting the equipment.
2 Specifications
• Power: 120VAC, 50/60Hz, 4.50A OR 230VAC, 50/60Hz, 2.30A
• Fuse:
o 120VAC, T5.0AL, 250VAC (5x20mm) Time delay low breaking capacity fuse
o 230VAC, T3.15AL, 250VAC (5x20mm) Time delay low breaking capacity fuse
• Temperature Scale: Celsius or Fahrenheit – user selectable
• Basin Temperature Set Point: 20° to 49° ± 1°C (68° to 120° ± 2°F)
• Use with single-use Ecolab Scope Warmer Pillow Drape (EPD400/EPD400N)
• Minimum fluid volume: 1.0L
• Maximum fluid volume: 5.5L
• Maximum working fluid volume: 4.5L
• Equipment mass including working load (unit, cord, fluid, scopes): 18.958 Kg
• Equipment mass unloaded (unit, cord): 6.658 Kg
• Adjustable Pole and Base mass (EFW-ABA): 6.03 Kg
• Non-adjustable Pole and Base mass (EFW-FBA): 5.33 Kg
• The device provides no protection against water.
• The EFW5L has been evaluated with respect to electromagnetic compatibility, electric
shock, fire and mechanical hazards in accordance with EN lEC 60601-1,
ANSI/AAMI/ES 60601-1:2005/A2:2021