Ecolab EFW5L - 3 Operating Characteristics

Ecolab EFW5L
51 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 15 of 51
EFW5L_IFU REV NEW
EFW5L Unit
Model
Number
Description
EFW5L
Ecolab Fluid Warmer
Table 6: EFW5L Unit
Power Cord
Figure
Model
Voltage
Plug Type
Main Region Used
CTP25NA-B
120V
Type B (Hospital
Grade)
United States, Canada,
Columbia, Costa Rica,
Ecuador, Dominican Republic,
Mexico
EFW25NA-
B
120V
Type B (Hospital
Grade)
United States, Canada,
Columbia, Costa Rica,
Ecuador, Dominican Republic,
Mexico
CTP25UK-G
230V
Type G
United Kingdom, Hong Kong,
Ireland, Singapore, United
Arab Emirates, Saudi Arabia,
Bahrain, Kuwait, Qatar
Table 7: Power Cord
3 Operating Characteristics
Maximum fluid Set Point temperature: 49°C (120°F)
Minimum fluid Set Point temperature: 20°C (68°F)
Warmer Basin fluid volume for use: 1.0-4.5 L
Language supported: English
Wheel Locks: Locks the wheels to prevent unintended movement of EFW5L.
transmitter
W
150 kHz to 80 MHz
d=1.2√P
80 MHz to 800 MHz
d=1.2√P
800 MHz to 2.5 GHz
d=2.3√P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1At 80 MHz to 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
Table 5: Recommended Separation Distances