EFW5L_IFU REV NEW
7.5 Post Procedure and Storage Steps
Refer to Figure 9.
a. Shut down EFW5L by following the steps in section 7.2.
b. Remove residual fluid and dispose in accordance with hospital guidelines.
c. Lift drape off EFW5L and dispose in accordance with hospital guidelines.
d. Unplug male end of Power Cord (1) from wall outlet.
e. Store Power Cord on Power Cord Hook (2).
f. Allow EFW5L warmer basin to cool down completely (approx. 5 minutes) and follow
cleaning procedures as described in Section 9.1.
g. Unlock the wheels using the steps in Section 6.1. Using the Handle (3), move
EFW5L to a storage location.
7.6 Power Interruption
In the event of a power interruption:
a. The unit will NOT automatically restart. After main power is restored, power on the
unit by following the steps in section 7.1.
1
2
3
Figure 9: Storage Steps