Disconnect power cord from the wall outlet
to remove power from the unit for 10
seconds.
Then reconnect power cord and press the
on side of the Power Switch to restart the
warmer system. If malfunction still exists,
contact customer service.
displays “Attention!
Alarm 1: High.”
warmer system
exceeds Set Point
Temperature by 2°C
(4°F) or greater
Wait for the fluid to cool down to within
2°C (4°F) of the Set Point Temperature.
Turn off the unit by pressing the off side of
the Power Switch.
Allow the unit to cool for 5 minutes.
Press the on side of the Power Switch to
restart the unit. If malfunction still exists,
contact customer service.
adjust temperature,
Controller does not
adjust it.
adjustment function
has been locked
Refer to Section 6.6.3 to unlock
temperature adjustment.
Contact customer service.
do not secure EFW5L
Contact customer service.
EFW5L enclosure.
damaged
Remove the unit from service. Contact
customer service.
corrosion.
cleaned as frequently
as instructed.
Dry the basin with a lint free cloth and
isopropyl alcohol after each use and scrub
the basin with a non-abrasive cleanser,
per cleaning guidelines (Section
9).
displays unusual
fluctuations in the
Incorrect installation of
drape.
Ensure the drape is evenly distributed and
in full contact with the basin walls. If
malfunction still exists, contact customer