SA Series Ambulatory Blood Pressure Monitor User Manual Safety Guidance
- 7 -
2. Do not use the equipment in a dusty environment with bad ventilation or in the
presence of corrosive.
3. Make sure that there is no intense electromagnetic interference source around the
equipment, such as radio transmitters or mobile phones etc. Attention: large medical
electrical equipment such as electrosurgical equipment, radiological equipment and
magnetic resonance imaging equipment etc. is likely to bring electromagnetic
interference.
4. The device and accessories are to be disposed of according to local regulations after
their useful lives. Alternatively, they can be returned to the dealer or the manufacturer
for recycling or proper disposal. Batteries are hazardous waste. Do NOT dispose of
them together with house-hold garbage. At the end of their lives hand the batteries
over to the applicable collection points for the recycling of waste batteries. For more
detailed information about recycling of this product or battery, please contact your
local Civic Office, or the shop where you purchased the product.
5. The time required for the ABPM to warm from the minimum storage temperature
between uses until it is ready for intended use is at least 2 hours; and the time
required for the ABPM to cool from the maximum storage temperature between uses
until it is ready for intended use is at least 2 hours.
6. The monitor should not be modified in any way.
7. Federal (U.S.) law restricts this device to sale by or on the order of a physician.
1.2.5 Contraindications
Do not use on patients with erratic, accelerated or mechanically controlled irregular heart rhythms,
including patient with arrhythmias.
Due to the strangulation risk posed by the cuff and hose, the SA series ABPM must not be within
reach of unsupervised children, and must not be used on unsupervised patients with limited
cognitive abilities.
The SA series ABPM is not intended for alarm triggering monitoring purposes in intensive care
units, and must not be used for blood pressure monitoring purposes in intensive care units or
during surgery.