SA Series Ambulatory Blood Pressure Monitor User Manual EMC Information
- 49 -
Appendix 2 EMC Information
Electromagnetic Emissions
Guidance and manufacture’s declaration - electromagnetic emission
The SA series device is intended for use in the electromagnetic environment specified below.
The customer or the user of the device should assure that it is used in such an environment.
Electromagnetic environment - guidance
The SA series device uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The SA series device is suitable for use in all
establishments, other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC/EN 61000-3-2
Voltage fluctuations/
flicker emissions
IEC/EN 61000-3-3
Electromagnetic Immunity
Guidance and manufacture’s declaration - electromagnetic immunity
The SA series device is intended for use in the electromagnetic environment specified below.
The customer or the user of the SA SERIES DEVICE should assure that it is used in such an
environment.
Electromagnetic
environment - guidance
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least 30%.