SA Series Ambulatory Blood Pressure Monitor User Manual EMC Information
- 50 -
Electrical fast
transient/burst
IEC/EN 61000-4-4
2 kV for power
supply lines
1kV for input/output
lines
Not applicable
Not applicable
Surge
IEC/EN 61000-4-5
1 kV line to line
2 kV line to ground
Not applicable
Not applicable
Power frequency
(50Hz/60Hz)
magnetic field
IEC/EN 61000-4-8
30A/m
30A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC/EN
61000-4-11
0 % U
T;
0.5 cycle
At , 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30
cycles )
Single phase: at 0°
0 % U
T
; 250/300
cycle
Not applicable
Not applicable
NOTE U
T
is the a.c. mains voltage prior to application of the test level.