SA Series Ambulatory Blood Pressure Monitor User Manual Cleaning, Care and Maintenance
- 37 -
Chapter 11 Cleaning, Care and Maintenance
Use only the EDAN-approved substances and methods listed in this chapter to clean or disinfect
your equipment. Warranty does not cover damage caused by using unapproved substances or
methods.
Edan Instruments has validated the cleaning and disinfection instructions provided in this User
Manual. It is the responsibility of the healthcare professional to ensure that the instructions are
followed so as to ensure adequate cleaning and disinfection.
11.1 General Points
Keep your device and accessories free of dust and dirt. To prevent the device from damage,
please follow the instructions:
Use only the recommended cleaning agents and disinfectants listed in this manual. Others
may cause damage (not covered by warranty), reduce product lifetime or cause safety
hazards.
Always dilute according to the manufacturer's instructions.
Unless otherwise specified, do not immerse any part of the equipment or any accessories in
liquid.
Do not pour liquid onto the equipment.
Do not allow liquid to enter the case.
Never use abrasive material (such as steel wool or silver polish).
Inspect the monitor and reusable accessories after they are cleaned and disinfected.
CAUTION
1. If you spill liquid on the monitor, or it is accidentally immersed in water, do not use;
contact your service personnel or EDAN service engineer.
2. Clean and disinfect the device and reusable accessories when the patient changes.
3. The equipment is chemically resistant to most cleaning agents, disinfectants and
non-caustic detergents used in hospital, but cleaning agents or disinfectants that are
not listed in this manual are not recommended. For example, didecyl dimethyl
ammonium bromide, which contains quaternary ammonium salt, may corrode the
equipment and accessories.
11.2 Cleaning