SA Series Ambulatory Blood Pressure Monitor User Manual Technical Specifications
- 47 -
A1.2 Environment Specifications
-20ºC (-4ºF)~+55ºC (+131ºF)
+5ºC (+41ºF) ~ +40ºC (+104ºF)
15%RH~95%RH
Non-Condensing
15%RH~95%RH
Non-Condensing
A1.3 Physical Specifications
106.5 mm × 62 mm × 27 mm, ±1 mm
160g ± 10g (excluding battery and cuff)
A1.4 Power Supply Specifications
Two AA alkaline batteries
A1.5 Performance Specifications
0 to 300 mmHg (0 to 40 kPa), ±3 mmHg (±0.4 kPa)
Measurement pressure range
Systolic: 25 to 290 mmHg (3.3 to 38.6 kPa)
Diastolic: 10 to 250 mmHg (1.3 to 33.2 kPa)
Mean error: ±5 mmHg (±0.67 kPa)
Standard deviation: Max. 8 mmHg (1.07 kPa)
40 to 240 bpm, accurate within ±3 bpm or ±3.5%,
whichever is greater.