SA Series Ambulatory Blood Pressure Monitor User Manual Cleaning, Care and Maintenance
- 43 -
1) PC and Monitor
♦ Avoid excessive temperature, sunshine, humidity and dirt.
♦ Put the dustproof coat on the monitor after use and prevent shaking it violently when
moving it to another place.
♦ Prevent any liquid from seeping into the equipment; otherwise the safety and the
performance of the monitor cannot be guaranteed.
2) Hose
♦ Integrity of the pneumatic hose should be checked regularly. Make sure that it is
conductible.
♦ Do not drag or twist the hose with excessive stress while using it...
♦ Align the hose to avoid twisting, knotting or crooking in a closed angle while using it.
♦ Once damage or aging of the hose is found, replace it with a new one immediately.
While usage will have an impact, it is expected the monitor will be in service for 10 years.
Replace the hose, cuff sleeve and other accessories according to your actual use. It is
recommended that you replace them once every year.
EDAN will make available on request circuit diagrams, component part lists, descriptions,
calibration instructions, or other information that will assist service personnel to repair those parts
of the equipment that are designated by EDAN as repairable by service personnel.
CAUTION
The device and accessories are to be disposed of according to local regulations after
their useful lives. Alternatively, they can be returned to the dealer or the manufacturer for
recycling or proper disposal.