4. Installaon
4.1 Safety
WARNING:
Ensure that the installaon technician is familiar with the safety procedures which
relate to the pump oil and the products handled by the pumping system.
WARNING:
We recommend that a hydrocarbon-prepared RV pump is not used for pumping
hazardous substances. PFPE-prepared pumps are suitable for oxygen applicaons:
refer to PFPE-prepared RV pump refer to PFPE-prepared RV pumps on page 56.
Obey the safety instrucons in this Secon and take note of appropriate precauons.
If not, injury to people and damage to equipment can result.
Prevent any part of the human body from coming into contact with the vacuum.
Ensure that the RV pump is suitable for the applicaon. If there is any doubt as to the
suitability of the RV pump for the applicaon, refer our guidelines on vacuum pump and
vacuum system safety or contact us for advice.
A suitably trained and supervised technician must install the RV pump. Obey the safety
instrucons listed below when installing the pump, especially when connecng the
pump into an exisng system. Details of specic safety precauons are given at the
appropriate point in the instrucons.
Wear the appropriate safety-clothing when coming into contact with contaminated
components is ancipated. Dismantle and clean contaminated components inside
a fume cupboard.
Vent and purge the vacuum system before starng installaon work.
Take suitable precauons to avoid the inhalaon of oil mist and excessive skin
contact with pump-oil, as prolonged exposure can be harmful.
Disconnect the other components in the pumping system from the electrical
supply so that they cannot be operated accidentally.
Safely route any electrical supply cables to prevent a trip hazard.
4.2 System design consideraons
Consider the following points when designing the pumping system:
We recommend the use of a foreline vacuum isolaon valve to allow the pump to
warm up before pumping condensable vapours or if a vacuum need to be
maintained when the pump is not running.
Avoid high levels of heat input to the pump from the process gases, otherwise the
pump may overheat and seize, and cause the motor thermal overload device to
open.
03/2021 - ©Edwards Limited
Page 25A65201880_AA
A65201880_AA - Installaon

Table of Contents

Other manuals for Edwards RV3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards RV3 and is the answer not in the manual?

Edwards RV3 Specifications

General IconGeneral
BrandEdwards
ModelRV3
CategoryWater Pump
LanguageEnglish

Related product manuals