5.3 Start-up procedure
WARNING:
Ensure that the system design does not allow the exhaust pipeline to be blocked.
If the oil is contaminated, or if the pump temperature is below 12 °C, or if the electrical
supply voltage is more than 10% below the lowest voltage specied on the voltage
indicator (Figure: Motor voltage conguraon: single-phase pumps, item 3), the pump
may operate at a reduced speed for a few minutes. On single-phase pumps, if the pump
connues to operate at reduced speed, the motor thermal overload device will open
and stop the pump. When the motor has cooled, the thermal overload device will reset
automacally and the pump will restart.
1.
Check that the pump oil-level is between the MAX and MIN marks on the bezel of
the oil-level sight-glass; if it is not, refer to Check the oil-level on page 41.
2.
Turn the mode selector fully clockwise to select High Vacuum mode S or fully
anclockwise to select High Throughput mode S, as required (refer to Check the
oil-level on page 41).
3.
Turn the gas-ballast control to posion ‘0’, ‘I’ or ‘II’, as required (refer to Gas-ballast
control on page 36).
4. Switch on the electrical supply to the pump; on single-phase pumps, use the on/o
switch.
5. In order to achieve ulmate vacuum, to pump condensable vapours or to
decontaminate the pump oil, refer to the procedures in To achieve ulmate
vacuum on page 37, To pump condensable vapours on page 38 and To
decontaminate the oil on page 38 respecvely. Otherwise, open the vacuum
system isolaon-valve.
5.4 To achieve ulmate vacuum
If the pump does not achieve the performance specied in Performance on page 13,
make sure that this is not due to the system design before contacng us or the supplier
for advice. In parcular, the vapour pressure of all materials used in the vacuum system
(including pump oil, see below) must be much lower than the specied ulmate vacuum
of the pump. Refer to The pump has failed to achieve the specied performance (has
failed to reach ulmate vacuum) on page 47 for a list of possible causes for failure to
achieve the specied performance. Note however that the most common causes are:
The pressure measurement technique or gauge head is unsuitable or the gauge
head is faulty.
If an oil other than the recommended oil has been used, and the vapour pressure
of the oil is higher than the specied ulmate vacuum of the pump.
Use the following procedure to achieve ulmate vacuum:
1.
Isolate the RV pump from the vacuum system.
2.
Turn the mode selector to select High Throughput mode S, set the gas-ballast
control to low ow (posion ‘I’) and operate the pump for at least 1 hour (or
overnight) to thoroughly purge the oil of contaminants.
03/2021 - ©Edwards Limited
Page 37A65201880_AA
A65201880_AA - Operaon

Table of Contents

Other manuals for Edwards RV3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Edwards RV3 and is the answer not in the manual?

Edwards RV3 Specifications

General IconGeneral
BrandEdwards
ModelRV3
CategoryWater Pump
LanguageEnglish

Related product manuals