13 Drain the unit Elma Schmidbauer GmbH
26 / 32 Elmasonic Select
12.12 Remove cleaned objects
ü The end of ultrasonic cleaning is indicated by the End of cleaning symbol or, where applicable, an
acoustic signal.
1. Carefully remove the lid, and allow any dripping water to drip into the tank. Then, place the lid
upside down next to the unit to use as a drip tray.
CAUTION!If using high cleaning temperatures > 50 °C, allow the fluid to cool first or use
suitable heat-resistant protective gloves.
2. Lift the basket or insert out of the tank, allow the excess fluid to drain briefly, and place it on the
upturned lid.
ðThe objects are now clean. Check the cleaning result.
è Check the fill level if you intend to start a further ultrasonic cleaning process.
è Switch off the unit. If the fluid is dirty or will not be used again, empty out the liquid. See the section
Drain the unit for information.
13 Drain the unit
CAUTION
Hot liquids
Danger of burns and scalds caused by spilled fluid!
Ø Do not move or carry the unit when full.
Ø Exception: unit sizes Elmasonic Select 30 to 60. Allow the fluid to cool beforehand,
and disconnect the mains plug.
Empty units with drains:
ü Ultrasound operation has finished.
ü The basket or insert has been removed.
ü The drain (1) is connected to an on-site wastewater disposal system.
1. Open the discharge tap (2) to drain off the cleaning fluid.
ðThe tank is empty.
2. Clean the tank and disinfect it if necessary. See chapter Maintenance for information.
CAUTION!Never submerge the unit in water.
è The unit is emptied, cleaned and, where necessary, disinfected.
1
1
2 2
Illustration6: Drain of Elmasonic Select 60–300/500/900