17 Disposal Elma Schmidbauer GmbH
32 / 32 Elmasonic Select
2. The blower will start after >50°C is reached.
ðIf the blower is defective, the sound made by it will be clearly audible or the display will show the
error message "Er47".
3. Press the On/off button to switch off the unit.
The blower function has been checked. If no sound can be heard from the blower or the error message
is shown, send the unit for repair immediately.
16.6 LED test
Test the LED function (see Controls [}16])
Regularly check the function of all LEDs on the display to ensure reliable operation. The display showing
all LED indicators is illustrated in the chapter Controls [}16]. Send the unit for repair immediately if any
LED indicators are not displayed.
The unit has been switched off.
Hold down the Mode button until the Setup menu is shown on the display.
To set the LED test function, press the Plus/minus buttons repeatedly until the
parameter number10 is shown.
Then, press the Plus button to change the parameter value from 00 to 01.
The LEDs will light up for approx. 5seconds. The parameter value will be reset
automatically. Repeat this procedure as required.
Press the On/off button to end the LED test.
17 Disposal
CAUTION
Once the unit has reached the end of its service life, ensure that the unit and
accessories are disposed of safely and correctly:
Ø Clean and disinfect the old device and accessories before disposal.
Ø Secure the old device against unauthorised access until removal; if necessary,
dispose of the power cable separately.
Ø Observe regionally applicable disposal directives.