Flow Measurement System Version 1.00 page 37 of 75
To calculate an output current into the “original” flow value as the system is measuring it
(without adjusted calibration factor), use the following equation:
Factorn Calibratio
range) flow(max Factor Conversion
4[mA])-[mA](Current n]Flow[ml/mi
The conversion factors are listed below:
Clamp-On
Transducer
Size IDxWT
Standard . flow
range [ml/min]for a
calibration factor of
1
Max. flow range
[ml/min] for
max calibration
factor 1.5
Conversion Factor
[ml/(min*mA)]
CT 11/64” x 3/64
6000
9000
562.5
CT 3/16 x 1/16”
6000
9000
562.5
CT 1/4 x 1/16”
8000
12000
750
CT 1/4 x 3/32”
8000
12000
750
CT 3/8 x 1/16”
10000
15000
937.5
CT 3/8 x 3/32”
10000
15000
937.5
CT 1/2 x 3/32”
20000
30000
1875
CT 1/2 x 1/8
20000
30000
1875
Please note: the conversion factor is a fixed value based on the maximum flow range of a
sensor when the calibration factor is set to 1.50. It is not dependant on the actual chosen
calibration factor.