Flow Measurement System Version 1.00 page 37 of 75
To calculate an output current into the “original” flow value as the system is measuring it
(without adjusted calibration factor), use the following equation:
Factorn Calibratio
range) flow(max Factor Conversion
4[mA])-[mA](Current n]Flow[ml/mi
The conversion factors are listed below:
Clamp-On
Transducer
Size IDxWT
Standard . flow
range [ml/min]for a
calibration factor of
1
Max. flow range
[ml/min] for
max calibration
factor 1.5
Conversion Factor
[ml/(min*mA)]
Please note: the conversion factor is a fixed value based on the maximum flow range of a
sensor when the calibration factor is set to 1.50. It is not dependant on the actual chosen
calibration factor.