8.2 Available Error Codes/Information for Flow Meter
In case of an error, the flow meter will not allow any additional input or
change of screen. However, the device can be powered off and on again. If
the error continues to appear, the device needs to be serviced, see chapter
SUPPORT AND SERVICE.
Please see also the chapter TROUBLESHOOTING.
The following error information is provided via the different interfaces:
Description
Digital
Interface
RS 232 Status
Byte “ee”
Analogue
Interface
4-20mA Flow
Analogue
Interface
4-20mA RSS
Display
(BioProTT
FlowTrack plus
only - on
flow screen)
No error
00
Flow value
[in mA]
RSS value
[in mA]
Flow value
RSS value
FLASH error
during start up
test
01
0 mA
0 mA
Error page w/
01
FLASH error
during runtime
test
02
0 mA
0 mA
Error page w/
02
RAM error
during runtime
RAM test
08
0 mA
0 mA
Error page w/
08
Error in task
handling
0F
0 mA
0 mA
Error page w/
0F
EEPROM test
error (Sensor)
19
0 mA
0 mA
Error page w/
19
Board
temperature
error (> 75°C)
1A
0 mA
0 mA
Error page w/
1A
+15V check
error
1B
0 mA
0 mA
Error page w/
1B
-15V check
error
1C
0 mA
0 mA
Error page w/
1C
EEPROM CRC
error (Sensor)
21
0 mA
0 mA
Error page w/
21
EEPROM
factor inconsis-
tent (Sensor)
22
0 mA
0 mA
Error page w/
22
EEPROM of
the sensor not
consistent
(Sensor)
27
0 mA
0 mA
Error page w/
27