Flow Measurement System Version 1.00 page 45 of 75
7.7.3.6 Status and Error Information
The following information is displayed on the status and error screens.
Ultrasonic Flow Meter
em-tec GmbH
www.em-tec.com
SW: FT_GFX_EV2200
SN: 000001
ERROR
0x2A
Illegal sensor type
or sensor version
When the system is starting up, the message “Initializing
Device” is displayed on the screen.
When accessing the status information screen, SW version and
serial number of the BioProTT™ flow meter are displayed. An
example is shown here
This information is important for troubleshooting.
When an error is detected, an error screen is shown with the
respective error code and message. An example is shown here.
Please refer to chapter AVAILABLE ERROR
CODES/INFORMATION FOR FLOW METER for detailed
error information.
This information is important for troubleshooting
7.7.4 Custom Calibration Procedure
The BioProTT™ provides the possibility to do a custom calibration to account for different
media, tubing wall thickness and temperatures.
The customization can be performed by sending the command C followed by the
calibration factor (see chapter MEASUREMENT USING THE DIGITAL INTERFACE
(RS232) OF
THE DIFFERENT BIOPROTT™ FLOW METERS) via the serial interface. Additionally
the factor can be set via the display for the BioProTT FlowTrack plus (see chapter
SELECTION OF CALIBRATION FACTOR)