Flow Measurement System Version 1.00 page 51 of 75
Problem
Meaning / Solution
supplied via process control
See also chapter AVAILABLE ERROR
CODES/INFORMATION FOR FLOW METER
0.2-0.3 mA
A reading of 0.2 0.3 mA indicates a circuit error.
>0.3 <1 mA
Undefined output
Analogue Output
RSS Output
0 mA
A reading of 0 mA could indicate the following:
- broken circuit
- missing power supply for the current loop; usually power is
supplied via process control
See also chapter AVAILABLE ERROR
CODES/INFORMATION FOR FLOW METER
0.2-0.3 mA
A reading of 0.2 0.3 mA indicates a circuit error.
>0.3 <4 mA
Undefined output
The maximum flow range depends on the size of the Clamp-On Transducer, please check
in the Clamp-on Transducer datasheet whether the given flow range is suitable for your
application.