BioProTT
TM
Flow Measurement System Version 1.00 page 7 of 75
Risk to safety due
to defective
device, applied
parts or
accessories
A defective device, sensors/parts or accessories must not be used
and must be replaced immediately. This applies in particular if the
cable insulation is damaged or if parts have broken off or are bent.
The power supply unit may not be immersed into liquids. However,
in the event of the ingress of liquid, pull out the power plug
immediately. Stop using the device and send it to our back for
inspection; see chapter RETURNS / RMA.
Malfunctions,
inaccurate
measured value
due to high-
frequency devices
The BioProTT
TM
flow measurement system may be influenced by
radio frequency (RF) devices. This includes mobile RF
communication equipment. Use of a RF device in the vicinity of the
BioProTT
TM
flow measurement system may also cause its
accessories to malfunction which may, among other things, lead to
measured values being provided inaccurately or incorrectly.
Overheating of
the device &
power supply
Prevent the device and power supply from being covered and direct
heat or sun radiation.
Repairs to the device shall only be carried out by authorized service
personnel. Unauthorized repairs will void manufacturer’s warranty.
Faulty electrical
installation
Comply with the installation requirements; refer to the chapter
ELECTRICAL SAFETY AND ELECTROMAGNETIC
COMPATIBILITY.
Measuring
inaccuracy,
Risk to electrical
safety
The measuring accuracy and safety-relevant aspects should be
checked at least every 2 years. Please refer to chapter
RECOMMENDED PERIODIC CHECKS for details
Risk of
infection/environ-
mental
contamination
due to improper
disposal
The device and accessories must be disposed of in accordance with
the national specifications for waste electrical and electronic
devices.
In case microbial contamination may be present on the device after
being used, we recommend you dispose of these separately in
accordance with the applicable regulations.
In case microbial contaminated, device or accessories shall be
cleaned and disinfected before they are sent to the manufacturer.
Do not return a device/sensor if you are not absolutely certain that
all traces of hazardous substances have been removed, e.g.
substances which have penetrated crevices or diffused through
plastic. Refer to chapter RETURN INFORMATION AND
DECLARATION OF CONTAMINATION for additional
information.
For more application-specific warnings, refer to the corresponding chapters.
The manufacturer only considers itself to be responsible for the safety, reliability and
usability of the device if:
The device is used in accordance with the instructions for use.