Operating Manual V 1.06 126 / 143
Test setup
Warning
For magnetic field testing the power mains input at PF1 shall be
disconnected.; 230 V/16 A.
The voltage V is adjusted with a variac as long as the required antenna current is available and the related H field
is generated in the center of the magnetic field antenna.
The variac NX1-260-16may be controlled automatically via the test generator. Any equivalent variac available in
the lab can be used to control the current manually. Please take care that the variac has a sufficient current
capability.
For more detailed information the operator can require an additional manual especially for magnetic field
testing. This manual is part of the delivery of magnetic field testing accessories.
Test setup with MC 2630 for H-Fields up to 30A/m
Schematics Test setup
Schematics and Test setup with compact NX5, variac NX1-260-16, MC 2630 and MS 100N
Option for required magnetic field tests as per IEC 61000-4-8
- External variable motor transformer variac NX1-260-16
- External Magnetic field antenna MS 100N
- External current transformer MC 2630 to test 1, 3, 10 and 30A/m levels

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EM TEST compact NX5 Series and is the answer not in the manual?

EM TEST compact NX5 Series Specifications

General IconGeneral
BrandEM TEST
Modelcompact NX5 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals