EM TEST Compact NX5
Operating Manual V 1.06 72 / 143
6.4.
Power Fail Generator as per IEC 61000-4-11
EUT supply
Model NX5
Other models NX series
Channel PF1 and PF2
AC voltage/current
max. 300V/16 A
Up to max. 400 V/ 32 A
Mains frequency
50/60 Hz
DC voltage/current
max. 300 V/16 A
Up to max. 300 V/32 A
Inrush current
more than 500 A
Protection
Electronic fuse for continuous overcurrent / inrush currents
Electronic control of overheating
PF1 and PF2 are safe against short circuit
max. current channel PF1
nominal current continuous
max. current channel PF2
nominal current continuous
Trigger
Events trigger
AUTO, MAN, EXTERN
Repetition rate
0.01s 9’999 s Accuracy 0.05% + phase sync
Duration of events
0.01 ms 9’999 s Accuracy 0.05%
Events preselection
1 - 99'999
Counter
1 1000000
Synchronization
0 - 360° (16 500 Hz)
Asynchronous = 0° if a reference signal is connected to the Sync input.
Max Sync input voltage
Max. 300 V or same as EUT voltage specs.
Resolution
Measurements
DUT supply
AC/DC voltage in the LCD display, divider ratio 1:100 +/- 10%
BNC output MON U
Measurement of the EUT supply divider ratio 1:100 5%
BNC output MON I
Measurement of the EUT current and the inrush current
10 mV/A 5%, max. 1000 A
CRO TRIGGER
positive going flank
0-10V Control Output
0-10 Vdc for external voltage source
Test routines
Quick Start
Immediate start, all parameters adjustable during a running test
Standard test routines
as per IEC 61000-4-11 ac power ports
as per IEC 61000-4-29 dc power ports
as per IEC 61000-6-1 Generic
as per IEC 61000-6-2 Generic
Manual operated standard test routine
User test routines
Voltage variation, external variac control
Change phase angle A after n events by A
Change events duration td after n events by td
Inverse mode
Magnetic field test
test routine according to IEC 61000-4-8
test level 1, 3, 10 and 30 A/m with MC 2630 and a variac
test level 100, 300 and 1000 A/m with MC26100
Magnetic field tests per IEC 61000-4-8 and -9
The test routines for handling the magnetic field tests are included in the internal compact NX5 firmware. All functions
to control external options as voltage/current sources or magnetic field antennas are included. In addition, the
following hardware is required:
Option required for Magnetic Field Test per IEC 61000-4-9
- Magnetic field antenna (square 1mx1m coil MS 100)
- Adapter for connecting the square coil to the surge output.
Option required for Magnetic Field Test 50/60Hz per IEC 61000-4-8
- External transformer (variac NX1-260-16) and magnetic field antenna (square 1mx1m coil MS 100)
- External current transformer (MC2630) to test 1, 3, 10 and 30 A/m levels
- External current transformer (MC26100) to test level 100, 300 and 1000 A/m levels (short term).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EM TEST compact NX5 Series and is the answer not in the manual?

EM TEST compact NX5 Series Specifications

General IconGeneral
BrandEM TEST
Modelcompact NX5 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals