Option Description
Fix at Last
Measured Value
During the test, all outputs will report the last measured value of their assigned
process variable. The test will run for approximately 140 seconds.
Fix at Fault During the test, all outputs will go to their configured fault action. The test will run for
approximately 140 seconds.
The test starts immediately.
5. Wait for the test to complete.
Tip
At any time during the process, you can abort the test. If the outputs were fixed, they will return to
normal behavior.
• When SMV is licensed, results for the test are stored in the transmitter memory, along with results for tests
performed with any other tool. These results are also stored in the ProLink III database. You can view and
use these results in any tool-based trending and reporting functions.
• When SMV is not licensed, results from previous tests are saved in a database, but not accessible. A Pass/
Fail indication is all that is provided for the current test.
Postrequisites
View the results and take any appropriate actions.
Run an SMV test using ProLink III
Procedure
1. Read the Prerequisites in Use Smart Meter Verification (SMV) if you have not done so already.
2. Choose Device Tools > Diagnostics > Meter Verification > Run Test.
Tip
If Smart Meter Verification is licensed in the 4200 transmitter, you can also scroll to the bottom of the
ProLink III main screen and choose from the shortcut buttons: Smart Meter Verification Overview >
Meter Verification > Run Verification.
ProLink III automatically compares the contents of its SMV database to the SMV database on the
device, and uploads test data as required. You may need to wait for a few seconds until this process is
complete.
3. In the SMV Test Definition window, enter any desired information and click Next.
None of this information is required. It does not affect SMV processing.
ProLink III stores this information in the SMV database on the PC. It is not saved to the transmitter.
4. Select the desired output behavior.
Option
Description
Fix at Last
Measured Value
During the test, all outputs will report the last measured value of their assigned
process variable. The test will run for approximately 140 seconds.
Measurement support Configuration and Use Manual
April 2019 MMI-20048166
126 Micro Motion 4200 Transmitters