Option Description
Fix at Fault During the test, all outputs will go to their configured fault action. The test will run for
approximately 140 seconds.
5. Click Start and wait for the test to complete.
Tip
At any time during the process, you can abort the test. If the outputs were fixed, they will return to
normal behavior.
When SMV is licensed, results for the test are stored in the transmitter memory, along with results for tests
performed with any other tool. These results are also stored in the ProLink III database. You can view and
use these results in any tool-based trending and reporting functions.
When SMV is not licensed, results from previous tests are saved in a database, but not accessible. A Pass/
Fail indication is all that is provided for the current test.
Postrequisites
View the results and take any appropriate actions.
Run an SMV testing using the Field Communicator
Procedure
1. Read the Prerequisites in Use Smart Meter Verification (SMV) if you have not done so already.
2. Choose Service Tools > Maintenance > Routine Maintenance > SMV > Manual Verification > Start.
3. Select the desired output behavior.
Option
Description
Fix at Last
Measured Value
During the test, all outputs will report the last measured value of their assigned
process variable. The test will run for approximately 140 seconds.
Fix at Fault During the test, all outputs will go to their configured fault action. The test will run for
approximately 140 seconds.
The test starts immediately.
4. Wait for the test to complete.
Tip
If the 'Continuous Measurement' option is selected, you cannot abort the meter verification.
When SMV is licensed, results for the test are stored in the transmitter memory, along with results for tests
performed with any other tool. These results are also stored in the ProLink III database. You can view and
use these results in any tool-based trending and reporting functions.
When SMV is not licensed, results from previous tests are saved in a database, but not accessible. A Pass/
Fail indication is all that is provided for the current test.
Postrequisites
View the results and take any appropriate actions.
Configuration and Use Manual Measurement support
MMI-20048166 April 2019
Configuration and Use Manual 127