6. Fill the sensor with the D2 fluid and allow the sensor to achieve thermal equilibrium.
7. Perform the D2 calibration.
a) Choose D2 (Water).
b) Enter the density of your D2 fluid.
c) Choose Start Calibration.
d) Wait for the calibration to complete.
e) Choose Finished.
8. Open the shutoff valve.
Postrequisites
If you disabled LD Optimization before the calibration procedure, re-enable it.
12.6.2 Perform a D1 and D2 density calibration using ProLink III
Procedure
1. Read the Prerequisites in Perform a (standard) D1 and D2 density calibration if you have not already
done so.
2. See the following figure.
Measurement support Configuration and Use Manual
April 2019 MMI-20048166
142 Micro Motion 4200 Transmitters