2. Contact customer service.
15.2.40 [121] Extrapolation Alert (Concentration)
Alert
Process Aberration
Cause
The line density or line temperature is outside the range of the concentration matrix plus the configured
extrapolation limit.
Recommended actions
1. Ensure that process density and temperature are within the limits of the concentration matrix.
2. Verify the configuration of the concentration measurement application.
15.2.41 [123] Pressure Overrange (API referral)
Alert
Process Aberration
Cause
The line pressure is outside the range of the API table.
Recommended actions
1. Ensure that process pressure is within the range of the API table.
2. Verify the configuration of the API referral application and related parameters.
15.2.42 [131] Smart Meter Verification in Progress
Alert
Function Check in Progress
Cause
A Smart Meter Verification test is running.
Recommended actions
Allow the test to complete.
15.2.43 [132] Sensor Simulation Active
Alert
Sensor Being Simulated
Cause
Sensor simulation mode is enabled (enhanced core processor only).
Configuration and Use Manual Troubleshooting
MMI-20048166 April 2019
Configuration and Use Manual 171