Recommended actions
Disable sensor simulation.
15.2.44 Clock Failure
Alert
Data Loss Possible
Cause
The transmitter's real-time clock is not incrementing.
Recommended actions
Contact customer service.
15.2.45 Enhanced Event X Active
Alert
Event Active
Cause
The conditions assigned to any enhanced event (Enhanced Event 1 through Enhanced Event 5) are present.
Recommended actions
1. If this is an accurate indication of process conditions, no action is required. The alert will clear when the
process returns to normal.
2. Review event configuration if you believe the event was triggered erroneously.
15.2.46 Moderate Two Phase Flow
Alert
Process Aberration
Cause
The transmitter has detected moderate two-phase flow.
Recommended actions
Check your process conditions against the values reported by the device.
15.2.47 No Permanent License
Alert
Data Loss Possible
Cause
A permanent license has not been installed in the transmitter firmware.
Troubleshooting Configuration and Use Manual
April 2019 MMI-20048166
172 Micro Motion 4200 Transmitters