Table 7-2: Options for alerts and conditions (continued)
Alert Conditions
Name Description Ignorable
Default severity: Function Check
Severity configurable: No
Fault Timeout applicable: No
[131] Meter Verification in
Progress
A meter verification test is running. Yes
Sensor Being Simulated
Default severity: Function Check
Severity configurable: No
Fault Timeout applicable: No
[132] Sensor Simulation
Active
Sensor simulation mode is enabled.
(Enhanced core processor only)
No
Output Fixed
Default severity: Function Check
Severity configurable: No
Fault Timeout applicable: No
[101] mA Output 1 Fixed The HART address is set to a non-
zero value, a loop test is running, or
the output is configured to send a
constant value (mA Output Action
or Loop Current Mode).
Yes
[114] mA Output 2 Fixed The output is configured to send a
constant value. A loop test may be in
progress.
No
[111] Frequency Output 1
Fixed
The output is configured to send a
constant value. A loop test may be in
progress.
No
[118] Discrete Output 1 Fixed The output is set to a constant state.
A loop test may be in progress.
No
Drive Over-Range
Default severity: Maintenance
Required
Severity configurable: Yes
Fault Timeout applicable: Yes
[102] Drive Overrange The drive power (current/voltage) is
at its maximum.
Yes
Process Aberration
Default severity: Out of
Specification
Severity configurable: Yes
Fault Timeout applicable: Yes
[105] Two-Phase Flow The line density is outside the user-
defined two-phase flow limits.
Yes
[115] External Input Error The connection to an external
measurement device has failed. No
external data is available.
Yes
[121] Extrapolation Alert
(Concentration)
The line density or line temperature
is outside the range of the
concentration matrix plus the
configured extrapolation limit.
Yes
[116] Temperature Overrange
(API referral)
The line temperature is outside the
range of the API table.
Yes
[117] Density Overrange (API
referral)
The line density is outside the range
of the API table.
Yes
[123] Pressure Overrange (API
referral)
The line pressure is outside the
range of the API table.
Yes
Configuration and Use Manual Configure device options and preferences
MMI-20048166 April 2019
Configuration and Use Manual 91