Table 7-2: Options for alerts and conditions (continued)
Alert Conditions
Name Description Ignorable
Moderate Two-Phase Flow The transmitter has detected
moderate two-phase flow.
Yes
Severe Two-Phase Flow The transmitter has detected severe
two-phase flow.
Yes
Event Active
Default severity: Out of
Specification
Severity configurable: Yes
Fault Timeout applicable: Yes
Enhanced Event 1 Active The conditions assigned to
Enhanced Event 1 are present.
Yes
Enhanced Event 2 Active The conditions assigned to
Enhanced Event 1 are present.
Yes
Enhanced Event 2 Active The conditions assigned to
Enhanced Event 2 are present.
Yes
Enhanced Event 3 Active The conditions assigned to
Enhanced Event 3 are present.
Yes
Enhanced Event 4 Active The conditions assigned to
Enhanced Event 4 are present.
Yes
Enhanced Event 5 Active The conditions assigned to
Enhanced Event 5 are present.
Yes
Output Saturated
Default severity: Out of
Specification
Severity configurable: Yes
Fault Timeout applicable: No
[100] mA Output 1 Saturated The calculated output value is
outside the range of the output.
Yes
[113] mA Output 2 Saturated The calculated output value is
outside the range of the output.
Yes
[110] Frequency Output 1
Saturated
The calculated output value is
outside the range of the output.
Yes
Function Check Failed or Meter
Verification Aborted
Default severity: Maintenance
Required
Severity configurable: Yes
Fault Timeout applicable: No
[010] Calibration Failed The calibration failed. No
[034] Meter Verification
Failed
The meter verification test showed
that the sensor response was not
acceptably close to the baseline.
Yes
[035] Meter Verification
Aborted
The meter verification test did not
complete, possibly because it was
manually aborted or because
process conditions were too
unstable.
Yes
Configuration Warning
Default severity: Maintenance
Required
Severity configurable: Yes
Fault Timeout applicable: No
No Permanent License A permanent license has not been
installed in the transmitter firmware.
No
Clock Failure The transmitter's real-time clock is
not incrementing.
No
Transmitter Software Update
Failed
The transmitter software update
failed.
Yes
Configure device options and preferences Configuration and Use Manual
April 2019 MMI-20048166