b) Make sure the process/calibration selector valve (B) remains on the calibration position.
c) As the gas flows, select the Start button and observe the Sensor Time Period and Stability
values.
d) Wait between 3 to 15 minutes for the system to stabilize. When Stability is Good, select Accept
or Next.
If measurement does not stabilize after 30 minutes, select Abort and troubleshoot the problem.
e) Disconnect the calibration gas.
10. Select Next.
11. Review the results for this calibration.
12. Select Finish to save the results and exit, or select Add Calibration to add a calibration.
Tip
You can use Add Calibration to create a new set of calibration coefficients.
13. Set Active Calibration to the calibration to be used for measurement.
Purging and calibration Configuration and Use Manual
March 2021 MMI-20020954
40 Micro Motion Gas Specific Gravity Meters (SGM)