Status alerts and options for Status Alert Severity
Alert number Alert title Default severity User can reset
severity
A001 EEPROM Error Fault No
A002 RAM Error Fault No
A003 No Sensor Response Fault Yes
A004 Temperature Overrange Fault No
A006 Characterization Required Fault Yes
A008 Density Overrange Fault Yes
A009 Transmitter Initializing/Warming Up Ignore Yes
A010 Calibration Failure Fault No
A014 Transmitter Failure Fault No
A016 Sensor Temperature (RTD) Failure Fault Yes
A020 Calibration Factors Missing Fault Yes
A021 Transmitter/Sensor/Software Mismatch Fault No
A029 Internal Electronics Failure Fault No
A030 Incorrect Board Type Fault No
A033 Insufficient Pickoff Signal Fault Yes
A037 Sensor Check Failed Fault Yes
A038 Time Period Signal Out of Range Fault No
A100 mA Output 1 Saturated Informational To Informational or
Ignore only
A101 mA Output 1 Fixed Informational To Informational or
Ignore only
A102 Drive Overrange Informational Yes
A104 Calibration in Progress Informational To Informational or
Ignore only
A106 Burst Mode Enabled Informational To Informational or
Ignore only
A107 Power Reset Occurred Informational Yes
A113 mA Output 2 Saturated Informational To Informational or
Ignore only
A114 mA Output 2 Fixed Informational To Informational or
Ignore only
A115 No External Input or Polled Data Informational To Informational or
Ignore only
A118 Discrete Output 1 Fixed Informational To Informational or
Ignore only
Configure device options and preferences Configuration and Use Manual