8.8 Europe
8.8.1 I1 ATEX Intrinsic Safety
Certificate:
Baseefa07ATEX0239X
Standards:
EN IEC 60079-0: 2018, EN 60079-11: 2012
Markings:
II 1 G Ex ia IIC T4 Ga, T4(-60 °C ≤ T
a
≤ +70 °C)
Ex ia IIC T4 Ga, T4(-60 °C ≤ T
a
≤ +40 °C)
For use with Rosemount SmartPower
™
power module part number
753-9220-0001, or for use with Emerson SmartPower option 701PBKKF, or
MHM-89004 Blue Power Module.
Sensor terminal
parameters (option code
32)
Fuel sensor terminal
parameters (option code
61)
Plunger arrival
transmitter parameters
(option code 52)
U
O
= 6.15 V U
O
= 7.8 V U
O
= 6.6 V
I
O
= 13.37 mA I
O
= 92 mA I
O
= 125 mA
P
O
= 21.76 mW P
O
= 180 mW P
O
= 202 mW
C
i
= 0.216 µF C
i
= 10 nF C
i
= 8.36 nf
C
OIIC
= 23.78 µF C
OIIC
= 9.2 µF L
i
= 0
C
OIIB
= 549.78 µF C
OIIB
= 129 µF C
o
= 74 nF
L
i
= 0 L
i
= 0 L
o
= 1.5 mH
L
OIIC
= 200 mH L
OIIC
= 4.2 mH N/A
L
OIIB
= 800 mH L
OIIB
= 16.8 mH N/A
L
OIIA
= 1000 mH L
OIIA
= 33.6 mH N/A
Special Condition for Safe Use (X):
1. The surface resistivity of the antenna is greater than 1 GΩ. To avoid
electrostatic charge build-up, it must not be rubbed or cleaned with
solvents or a dry cloth.
2. The Model 701PBKKF Power Module, MHM-89004 Blue Power
Module or Intelligent Power Module 71008 may be replaced in a
hazardous area. The Power Modules have a surface resistivity greater
than 1 GΩ and must be properly installed in the wireless device
enclosure. Care must be taken during transportation to and from the
point of installation to prevent electrostatic charge build-up.
3. The Model 702 enclosure may be made from aluminum allow and
given a protective polyurethane paint finish; however, care should be
Quick Start Guide April 2021