10. COMPATIBILITY WITH DOLORCLAST
®
PRODUCT RANGE
The Dolorclast
®
Radial Shock Wave device and the BLUE handpiece are designed to be used together. The BLUE
handpiece can not be used with legacy Dolorclast
®
systems.
11. PRODUCT STORAGE AND SHIPPING
Storage and transport conditions are specied in
the “Technical Data” section.
Keep the original packaging until the product is to be
disposed of permanently. You can use it for shipping
and storage.
Should you wish to store or ship your product:
Proceed as described in the “Cleaning and Disin-
fecting” section,
Place the product and all accessories in the original
packaging.
12. ELECTROMAGNETIC COMPATIBILITY
The DOLORCLAST
®
Radial Shock Waves should not be used adjacent to or stacked with another unit. If
adjacent or stacked use is necessary, the DOLORCLAST
®
Radial Shock Waves should be observed to verify
normal operation in the conguration in which it will be used.
Portable RF communications equipment (including peripherals such as antenna cable and external antennas)
should be used no closer than 30 cm (12 inches) to any part of the DOLORCLAST
®
Radial Shock Waves, inclu-
ding cables specied by the manufacturer. Otherwise, degradation of the performance of this equipment could
result. The DOLORCLAST
®
Radial Shock Waves uses RFID transmission (13.56MHz) to communicate with
piston.
Instructions for maintaining BASIC Safety and Essential performance for the expected Service life
Do not change the once installed nal application due to EM DISTURBANCE. If the environment doesn’t correspond to
the conditions listed by the manufacturer, some actions are required to match those conditions. Please contact the
manufacturer.
The climatic environmental conditions could affect the life of critical components of the DOLORCLAST
®
Radial Shock
Waves.
The presence of transmitters near the DOLORCLAST
®
Radial Shock Waves could affect its performances. The
distances mentioned in the tables prepared by manufacturer could help to prevent any disturbances of the equipment
in normal operation.
00_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 3300_FB-656_10_rev_A_IFU_Dolorclast RSW.indb 33 25/03/2021 17:0725/03/2021 17:07

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EMS DOLORCLAST and is the answer not in the manual?

EMS DOLORCLAST Specifications

General IconGeneral
BrandEMS
ModelDOLORCLAST
CategoryMedical Equipment
LanguageEnglish

Related product manuals