Operation SPECORD PLUS
26
Tab Input
¡ Click on [
Add samples
], define the reference in accordance with the
screenshot and confirm the entries with [
Ok
]:
¡ Click on [
Add samples
] again and define the sample:
¡ In the second half of the sample table, enter "Holmium" as the name
and confirm with the ENTER key.
Archive auto-
matically
Make no entries.
Note
Enter the following note: "Spectral scan example"
} Exit parameter entry by clicking on [
Ok
].
ü The
Spectrum - Settings
window closes.
Performing the measurement
} Click on [
Measure sequence
] in the document window.
The start information for measuring the reference is displayed.
} Insert a reference sample (filter F0) and click [
Yes
] to start measurement.
The reference measurement is performed. After the measurement is complete, a
prompt for measurement of the "Holmium" sample appears.
} Insert the "Holmium" sample (holmium glass filter) and click [
Yes
] to start measure-
ment.
ü The holmium spectrum appears in the document window.