Cerabar M, Deltabar M, Deltapilot M Commissioning with an operating menu (onsite display/FieldCare)
Endress+Hauser 117
Expert System Instrument info
Expert System Display
Parameter name Description
Pd-tag. (022)
Display
Physical device tag
Example:
Deltabar M: EH_Deltabar_M_5x_6B032A0109D
Serial number (096)
Display
Displays the serial number of the device (11 alphanumeric characters).
Firmware version (095)
Display
Displays the firmware version.
Ext. order code (097)
Display
Displays the extended order code (max. 60 alphanumeric characters).
Factory setting
As per order specifications
Order code (098)
Display
Displays the order code (max. 20 alphanumeric characters).
Factory setting
As per order specifications
ENP version (099)
Display
Displays the ENP version
(ENP = electronic nameplate)
Electr. serial no. (121)
Display
Displays the serial number of the main electronics (11 alphanumeric characters).
Sensor serial no. (122)
Display
Displays the serial number of the sensor (11 alphanumeric characters).
Parameter name Description
Language (000)
Options
Select the menu language for the onsite display.
Options:
•English
Possibly another language (as selected when ordering the device)
One further language (language of the manufacturing plant)
Factory setting:
English
Display mode (001)
Options
Specify the contents for the first line of the onsite display in the measuring mode.
Options:
Only primary value (value+bar graph)
External value only (value+status)
All alternating (primary value+secondary value+ext.value)
Factory setting:
Primary value (PV)
Add. disp. value (002)
Options
Specify the contents for the second line of the onsite display in the measuring
mode.
Options:
•No value
Pressure
Main value (%)
Totalizer 1 (Deltabar M)
Totalizer 2 (Deltabar M)
The options depend on the measuring mode chosen.
Factory setting:
No value

Table of Contents