Cerabar M, Deltabar M, Deltapilot M Troubleshooting
Endress+Hauser 223
M438 Data record 23 10 The supply voltage is disconnected when
writing.
An error occurred when writing.
1. Check setting
2. Restart the device
3. Replace electr.
M472 Buffer 17 6 Writing too often to EEPROM Reduce write accessing to
EEPROM.
M515 Configuration
Flow
18 13 Max. flow out of nominal range of
sensor
1. Recalibrate the device
2. Restart the device
M882 Input signal 22 0 External measured value displays a
warning status.
1. Check the bus
2. Check source device
3. Check the setting
S110 Working range T 20 8 Over temperature and low temperature
present.
Electromagnetic effects are greater than
specifications in the technical data.
Sensor defect.
1. Check proc. temp.
2. Check temperature range
S140 Working range P 20 8 Overpressure and low pressure present.
Electromagnetic effects are greater than
specifications in the technical data.
Sensor defect.
1. Check the process pressure
2. Check the sensor range
S822 Process temp. 17 8 The temperature measured in the sensor
is greater than the upper nominal
temperature of the sensor.
The temperature measured in the sensor
is lower than the lower nominal
temperature of the sensor.
1. Check the temperature
2. Check the setting
S841 Sensor range 17 8 Overpressure or low pressure present.
Sensor defect.
1. Check the pressure value
2. Contact Endress+Hauser Service