5.3 Damping
Output dampingThe damping influences the time it takes for the display in "Measured Value" to react to
a change in pressure.
# Parameter Entry Significance
1 Set Output
Damping
e.g. 30 Damping in
transducer block
increased to 30 s
Note!
Both the transducer block output and the analog input block can be damped. We
recommend that the damping be set in the transducer block during commissioning and
increased if necessary in the analog input block during normal operation. This prevents
the device switching to "Out of Service" when the parameter is edited.
Note!
0 τ 2τ 3τ
63 %
100 %
BA211Y41
pressure
signal
time
Cerabar S FF 5 Pressure Measurement
Endress+Hauser 35