7.2 Simulation
Simualtion of the
OUT value
The OUT value can also be simulated in the analog input block:
1. Check that the simulation is enabled at the DIP-switch WP, see Chapter 3.2.
2. Click on the SIMULATE parameter:
– Set to enable
– Enter a simulation value
– Click on WRITE CHANGES button
– Check the effect on the OUT value and any following links.
3. Disable the SIMULATE parameter.
During simulation, the BLOCK_ERROR shows the status "simulation active".
7.3 Reset
By entering a code, the entries in the manufacturer's-specific parameters are reset
completely to factory settings.
# Parameter Entry Significance
1 Default
Values
5140 Reset to factory
setting
The following table provides a summary of factory settings. (* = display value)
Parameter Default Value
Measured Value *
Lower Range Value
0.0
Upper Range Value
= High Sensor
Limit
Set Lower Range Value
Set Upper Range Value
Set Bias Pressure
0.0
Bias Pressure Automatically
Set Output Damping
0.0
Select Pressure Unit
bar
Diagnostic Code *
0
Last Diagnostic *
0
Clear Last Diagnostic Code
Transducer_sw_vers *
Min. Pressure *
Act. pressure
Reset Min. Pres Peak Hold
Max. Pressure *
Act. pressure
Reset Max. Pres Peak Hold
Integer Counter high *
0
Sensor Temperature *
Act. temperature
Min. Temperature *
Act. temperature
Reset min Temp. Peak Hold
Max. Temperature *
Act. temperature
Reset max Temp. Peak Hold
"Pressure linear"
Parameter
Default Value
Default Values
Linearization
(operating mode)
Start Point
0.0
Full Scale
100.0
Unit After Liniarization
bar
Density Factor
1.0
Creep Flow Suppression
0.0
Low Sensor Calibration
= Low Sensor Limit
High Sensor Calibration
= High Sensor
Limit
Low Sensor Limit *
High Sensor Limit *
Sensor Pressure *
Act. sensor press.
Temperature Unit
°C
Unbiased Pressure *
Act. sensor press.
Biased Pressure *
Act. sensor press.
Security Locking
130
Serial-No. SENS. *
Process Conn. P+
special
Process Conn. P–
special
Gasket
special
Process Diaphram.
special
Fill Liquid
special
7 Diagnosis and Trouble-Shooting Cerabar S FF
48 Endress+Hauser