Mounting Deltabar PMD75B PROFINET with Ethernet-APL
18 Endress+Hauser
Level measurement in a closed vessel
+
min.
2
1
3
max.
5
1
4
ESC
-
+ E
A0038217
1 Shutoff valves
2 Device
3 Three-valve manifold
4 Drain valves
5 Separator
Mount the device below the lower measuring connection so that the piping is always
filled with liquid
Always connect the low-pressure side above the maximum level
When measuring in media with solid parts, such as dirty liquids, installing separators
and drain valves is useful for capturing and removing sediment
Level measurement in a closed vessel with superimposed vapor
+
min.
5
3
4
max.
2
1
6
2
ESC
-
+ E
A0038216
1 Condensate trap
2 Shutoff valves
3 Device
4 Three-valve manifold
5 Drain valves
6 Separator