Deltabar PMD75B PROFINET with Ethernet-APL Mounting
Endress+Hauser 19
Mount the device below the lower measuring connection so that the piping is always
filled with liquid
Always connect the low-pressure side above the maximum level
The condensate trap ensures constant pressure on the low-pressure side
When measuring in media with solid parts, such as dirty liquids, installing separators
and drain valves is useful for capturing and removing sediment
5.2.3 Pressure measurement
Pressure measurement with 160 bar (2 400 psi) and 250 bar (3 750 psi) measuring
cell
+
1
2
3
4
ESC
-
+ E
A0038218
1 Device with blind flange on low-pressure side
2 Three-valve manifold
3 Shutoff valves
4 Pressurized container
Mount the device above the measuring point so that the condensate can drain into the
process pipe
The negative side is open to atmospheric pressure via the screwed-in reference air filter
of the side flange on the low-pressure side
5.2.4 Differential pressure measurement
Differential pressure measurement in gases and vapors
+
1
2
3
4
ESC
-
+ E
A0043423
1 Device
2 Three-valve manifold
3 Shutoff valves
4 e.g. filter