Commissioning Deltabar PMD75B PROFINET with Ethernet-APL
52 Endress+Hauser
9.7.4 Commissioning with the commissioning wizard
In the Web server, SmartBlue and on the display, the Commissioning wizard is available
to guide the user through the initial commissioning steps.
1. Connect the device to the Web server.
2. Open the device in the Web server.
The dashboard (homepage) of the device is displayed:
3. In the Guidance menu, click the Commissioning wizard to open the wizard.
4. Enter the appropriate value in each parameter or select the appropriate option. These
values are written directly to the device.
5. Click "Next" to go to the next page.
6. Once all the pages have been completed, click "End" to close the Commissioning
wizard.
If the Commissioning wizard is cancelled before all the necessary parameters have
been configured, the device may be in an undefined state. In such situations, it is
advisable to reset the device to the factory default settings.