Deltabar S
Endress+Hauser 35
Form for device configuration – Level*
Operation via: Handheld terminal FieldCare/DeviceCare Onsite display
Device designation: Serial number:
Measuring point: Upper range limit (URL Sensor):
Parameter name Display ID Group Factory
setting
1)
Permitted
settings
Specified
value
Read-out
actual
value
Checked
Digits Set 840 Display 01234567890.-
Calib. Offset 319 Position Adjustm. 0.0
2)
Measuring Mode 389 Pressure Level
Level Selection 020 Level Easy
Pressure
Level Easy Pres-
sure
Empty Calib. 010 Basic Setup 0.0%
3)
Empty Pressure 011 0.0
) )
Full Calib. 004 100.0%
)
Full Pressure 005 URL Sensor
) )
Set LRV 013 0.0%
)
Set URV 012 100.0%
)
Damping Value 247 2.0 s 0 to 999 s
Press Eng. Unit 060 mbar / bar
),4)
all, except
"User unit"
Output Unit 023 %
)
Adjust Density 007 Extended setup 1.0 kg/dm³ = Process Density
(025)
Curr. Charact. 695 Output Linear Linear
Output Fail Mode 388 Max. alarm Max. alarm
Min. alarm
Alt. Current Output 597 Normal Normal
Max. Alarm Current 342 22 mA 21 to 23 mA
Set Min. Current 343 3.8 mA 3.8 mA or
4.0 mA
Simulation 413 Simulation None None
*Not for options with measuring range, e.g. 160/250 bar [2320/3625 psi])