Cerabar S/Deltabar S/Deltapilot S FOUNDATION Fieldbus Level measurement (FF configuration program)
Endress+Hauser 15
5.2 "Level easy pressure" level selection
5.2.1 Calibration with reference pressure – wet calibration
Example:
In this example, the level in a tank should be measured in m. The maximum level is 3 m (9.8 ft).
The pressure range is set to 0-300 mbar (4.5 psi).
Prerequisite:
The measured variable is in direct proportion to the pressure.
The tank can be filled and emptied.
!
Note!
See also Operating Instructions for Deltabar S (BA00301P) or Cerabar S (BA00302P), "Level
measurement" section or Deltapilot S (BA00372P), "Level measurement" section.
The values entered for Empty Calib. Level Easy/EMPTY_CALIBRATION_EASY and Full Calib.
Level Easy/FULL_ CALIBRATION_EASY must be at least 1% apart for the "Level Easy Pressure"
level mode. The value will be rejected with a warning message if the values are too close together.
Further limit values are not checked; i.e. the values entered must be appropriate for the sensor
and the measuring task so that the measuring device can measure correctly.
Due to the orientation of the device, there may be a shift in the measured value, i.e. when the
container is empty, the MEASURED VALUE parameter does not display zero.
For information on how to perform position adjustment, see also ä 59,
"Pos. Zero Adjust/ZERO_POSITION_ ADJUST".
Description
1 Deltabar S: Before configuring the device for your
application, the pressure piping must be cleaned and the
device filled with fluid. See Operating Instructions
BA00301P.
P01-PMP75xxx-19-xx-xx-xx-008
Fig. 1: Calibration with reference pressure –
wet calibration
1 See Table, Step 8.
2 See Table, Step 10.
2 Open the Pressure Transducer Block and set the block
mode to OOS.
0 mbar
300 mbar
3 m
0 m