Table 32: OPERATING MENU OPERATION
Parameter name Description
ENTER RESET CODE (047)
Entry
Reset parameters completely or partially to factory values or delivery status.
See also Operating Instructions for Deltabar S (BA00301P), Cerabar S (BA00302P) or
Deltapilot S (BA00372P), "Factory setting" (reset) section.
Factory setting:
0
OPERATING HOURS (409)
Display
Displays the hours of operation. This parameter cannot be reset.
INSERT PIN NO (048)
Entry
For entering a code to lock or unlock operation.
!
Note!
The -symbol on the local operation indicates that operation is locked. Parameters
which refer to how the display appears, e.g. LANGUAGE and DISPLAY CONTRAST,
can still be altered.
If operation is locked by means of the DIP switch, you can only unlock operation again
by means of the DIP switch. If operation is locked by means of remote operation e.g.
FieldCare, you can only unlock operation again using remote operation.
See also Operating Instructions for Deltabar S (BA00301P), Cerabar S (BA00302P) or
Deltapilot S (BA00372P), "Locking/unlocking operation" section.
Options:
Lock: Enter the number 0.
Unlock: Enter the number 2457.
Factory setting:
2457
HistoROM AVAIL. (831)
Display
Indicates whether the optional HistoROM®/M-DAT memory module is connected to the
electronic insert.
See also Operating Instructions for Deltabar S (BA00301P), Cerabar S (BA00302P) or
Deltapilot S (BA00372P), "HistoROM
®
/M-DAT (optional)" section.
Options:
•Yes (HistoROM
®
/M-DAT is attached to the electronic insert)
No (HistoROM
®
/M-DAT is not attached to the electronic insert)
DOWNLOAD SELECT (014)
Selection
Select download function from HistoROM to device. The option selected has no effect on
an upload from the device to the HistoROM.
Prerequisite:
•A HistoROM
®
/M-DAT is attached to the electronic insert
(HistoROM AVAIL. = Yes)
Options:
Configuration copy: With this option, all parameters apart from the TRANSMITTER
SERIAL No, DEVICE DESIGNATION and the parameters of the POSITION
ADJUSTMENT and PROCESS CONNECTION group are overwritten.
Device replacement: With this option, all parameters except for TRANSMITTER
SERIAL No, DEVICE DESIGNATION and the parameters of the POSITION
ADJUSTMENT and PROCESS CONNECTION group are overwritten.
Electronics replace: With this option, all parameters except for the parameters of the
POSITION ADJUSTMENT group are overwritten.
Factory setting:
Configuration copy (if HistoROM
®
/M-DAT is attached to the electronic insert)
HistoROM CONTROL (832)
Selection
For selecting the direction for copying the data.
See also Operating Instructions for Deltabar S (BA00301P), Cerabar S (BA00302P) or
Deltapilot S (BA00372P), "HistoROM
®
/M-DAT (optional)" section.
Prerequisite:
•A HistoROM
®
/M-DAT is attached to the electronic insert
(HistoROM AVAIL. = Yes)
Options:
Abort
•HistoROM Device
Device HistoROM
Factory setting:
Abort (if HistoROM
®
/M-DAT is attached to the electronic insert)